Reported Nov 20, 2013 · Terminated
DePuy Mitek FMS Outflow Tubing with One-Way valve Product Code:284649 Product Usage: The Intermediary Tubing for FMS¿…
DePuy Mitek, Inc., a Johnson & Johnson Co.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: DePuy Mitek, Inc., a Johnson & Johnson Co. · 325 Paramount Drive · Raynham, MA · 02767-5199
Is this the product you have?
- DePuy Mitek FMS Outflow
- Tubing with One-Way valve Product Code284649
- Product UsageThe Intermediary Tubing for FMS¿ Fluid Management Systems connects the Inflow Tubing to the arthroscopic sheath during arthroscopic procedures. The Polyvinyl Chloride (PVC) tubing integrates
- several functional components includingluer locks for attaching the tube to the Inflow tubing and sheath, clamps, and a one-way valve. The FMS Outflow Tubing with One-Way valve is sterile and intended for single use with the FMS Fluid Management Systems in a surgical setting by personnel trained in arthroscopy
- Lot codes and dates
- All lot codes
- Amount recalled
- 29280 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
Why it was recalled
Pillow valve included in the FMS Duo outflow tubing set and the FMS Solo intermediary tubing set may not perform as intended. may result in backflow of irrigation fluid into the One Day Set which may potentially lead to patient cross-contamination
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new DePuy Mitek, Inc., a Johnson & Johnson Co. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.