Reported Dec 1, 2021 · Ongoing

Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition

GE Healthcare, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE Healthcare, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615

Is this the product you have?

  • Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
Lot codes and dates
  • Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
Amount recalled
105 in total

Where it was sold

  • Nationwide

Why it was recalled

The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new GE Healthcare, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:00 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.