Reported Nov 22, 2023 · Ongoing

IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured…

Ion Beam Applications S.A.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Ion Beam Applications S.A. · Chemin Du Cyclotron 3 · Ottignies-Louvain-La-Neuve, N/A · N/A

Is this the product you have?

  • IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured with Orion Patient Positioning System.
Lot codes and dates
  • UDI-DI(01)05404013801138
  • Serial numbers (US)
    • SAT133
    • SAT140
    • SBF113
    • SBF128
  • SBF135 & (OUS)
    • SAT125
    • SAT127
    • SAT132
    • SAT136
    • SBF107
    • SBF112
    • SBF117
    • SBF124
    • & SBF125
Amount recalled
5 (US), 9 (OUS)

Where it was sold

Why it was recalled

The patient positioning system can perform motions in unexpected directions, creating the potential for crushing of the patient between the patient positioning devices.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ion Beam Applications S.A. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:42 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.