Reported Jun 12, 2024 · Ongoing
IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of patients…
Ion Beam Applications S.A.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Ion Beam Applications S.A. · Chemin Du Cyclotron 3 · Ottignies-Louvain-La-Neuve, N/A · N/A
Is this the product you have?
- IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
- Lot codes and dates
- UDISAT140: (01)05404013814015(11)190507(21)SAT140
- SAT136(01)05404013801138(11)201105(21)SAT136
- Serial NumbersSAT136, SAT140 PTS-11 versions from PTS-11.1 and PTS-12 versions, with C230 accelerator
- Amount recalled
- 2 units
Where it was sold
- Nationwide
- VA
Why it was recalled
Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Ion Beam Applications S.A. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.