Reported Jan 17, 2018 · Terminated
Philips Healthcare Brilliance iCT SP computed tomography x-ray system
Philips Medical Systems (Cleveland) Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Medical Systems (Cleveland) Inc · 595 Miner Rd · Cleveland, OH · 44143-2131
Is this the product you have?
- Philips Healthcare Brilliance iCT SP computed tomography x-ray system
- Lot codes and dates
- Brilliance iCT SP - 728311 Devices manufactured 03/05/2009 to present.
- Amount recalled
- 134
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Why it was recalled
There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the gantry.