ECG Out Cable - Heartstart MRx M1783A/M5526A 12-pin Sync cables All units of Philips sync cables model M1783A and…
Philips Electronics North America Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Electronics North America Corporation · 3000 Minuteman Rd · Andover, MA · 01810-1032
Is this the product you have?
- ECG Out Cable - Heartstart MRx M1783A/M5526A 12-pin Sync cables All units of Philips sync cables model M1783A and M5526A as well as Philips ECG out cables model 989803195641. Manufactured - September 4, 2002 October 8, 2014
- Lot codes and dates
- M1783A/M5526A 12-pin Sync cables All units of Philips sync cables model M1783A and M5526A as well as Philips ECG out cables model 989803195641
- Amount recalled
- 15811
Why it was recalled
Issue with ECG out cables. When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable or ECG out cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast transients (EFTs): " On the HeartStart MRx and HeartStart XL, EFT noise may be misinterpreted as an R-wave. " On the HeartStart XL+, EFT noise can disable ECG monitoring and potentially interrupt demand mode pacing*. *Note: It is contrary to the XL+ Instructions for Use to perform demand mode pacing while using the ECG out cable or obtaining the ECG signal from a bedside monitor. The XL+ Instructions for Use includes the following warning: When pacing in Demand Mode, the ECG cable from the patient must be directly connected to the HeartStart XL+. If the user follows this warning, this problem cannot occur on the XL+.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips Electronics North America Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.