Reported Jan 24, 2018 · Terminated

Persona Partial Knee System, Size 8, Finishing Guide

Zimmer Biomet, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Zimmer Biomet, Inc. · 56 E Bell Dr · Warsaw, IN · 46582-6989

Is this the product you have?

  • Persona Partial Knee System, Size 8, Finishing Guide
Lot codes and dates
  • Item No. 42578100801, Lot No. 63518668, UDI (01)00880304809253(10)63518668. Item No. 42578100802, Lot No. 63518669, UDI (01)00880304809338(10)63518669.
Amount recalled
158 devices

Where it was sold

Why it was recalled

Finishing guide does not adequately account for shape differences present in Size 8 of Persona Partial Knee System, potentially resulting in additional surgery time or sensation of tightness during knee flexion.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Zimmer Biomet, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 2, 2019 · Data as of Sep 19, 6:48 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.