Reported Nov 27, 2013 · Terminated
BIOMET Spine, 5.5 TI 5.5x50mm Translation Screw Polaris 5.5 TI Spinal System. Rx Only. Non-Sterile. Intended for…
Ebi, LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Ebi, Llc · 100 Interpace Pkwy · Parsippany, NJ · 07054-1149
Is this the product you have?
- BIOMET Spine, 5.5 TI 5.5x50mm Translation Screw Polaris 5.5 TI Spinal System. Rx Only. Non-Sterile. Intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system
- Lot codes and dates
- Multiple Item and Lot numbers. The firm expects all Translation Screw Systems (containing a cobalt chrome component) used and contacted with the SpF Implantable Spine Fusion Stimulator (DC electrical current) to be at risk for this failure mode.
- Amount recalled
- 7,092 units
Where it was sold
- Nationwide
- PR
Why it was recalled
Polaris Translation Screw Failures were recently reported to Biomet Spine. These failures are generally described as dissociation of the screw head from the shaft of the Polaris Translation Screw.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new infant & baby recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.