Reported Dec 31, 2014 · Terminated

Solitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sleeve Intended…

Ebi, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Ebi, Llc · 399 Jefferson Rd · Parsippany, NJ · 07054-3707

Is this the product you have?

  • Solitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sleeve Intended to stabilize cervical spinal segment to promote fusion in order to restrict motion and decrease pain using bone graft with or without supplemental fixation.
Lot codes and dates
  • Part number 14-531593 Part number 14-531594 Catalog number Lot Number 14-531593 N23597 14-531594 N23599
Amount recalled
193 units

Where it was sold

  • Nationwide

Why it was recalled

A design deficiency was discovered whereby the Drill/Awl Sleeve and Spring-Loaded Drill/Awl Sleeve instruments (the Sleeves) allow the mating Drill and Awl instruments to penetrate 1.0 mm beyond their intended depths during screw hole preparation.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ebi, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 15, 2015 · Data as of Sep 19, 5:21 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.