Reported Dec 12, 2018 · Terminated

GE HEALTHCARE CARESCAPE R860 Inspiratory Safety Guard, Part Numbers: a) 2066713-001 (single pack) b) 2083208-001 (box…

GE Healthcare, LLC

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE Healthcare, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615

Is this the product you have?

  • GE HEALTHCARE CARESCAPE R860 Inspiratory Safety Guard
  • Part Numbersa) 2066713-001 (single pack) b) 2083208-001 (box of ten singles) The CARESCAPE R860 ventilator is designed to provide mechanical ventilation or support to neonatal, pediatric, and adult patients weighing 0.25 kg and above. The CARESCAPE R860 ventilator is a microprocessor based, electronically controlled, pneumatically driven ventilator that includes integrated monitoring of FiO2, airway pressure, flow, and volume
Lot codes and dates
  • a) 2066713-001 Lot Numbers
    • 17/00951
    • 17/01174
    • 17/01937
    • 17/01994
    • 17/02372
    • 17/02393
    • 18/00126
    • 18/00127
    • 18/00128
    • 18/00129
    • 18/00130
  • b) 2083208-001 Lot Numbers
    • 17/00951
    • 17/01174
    • 17/01937
    • 17/01994
    • 17/02372
    • 17/02393
    • 18/00126
    • 18/00127
    • 18/00128
    • 18/00129
    • 18/00130
Amount recalled
1140 units (307 units in US)

Where it was sold

  • Nationwide
  • PR

Why it was recalled

Stop using this product now

The ventilator Inspiratory Safety Guard (ISG) may disconnect from the breathing circuit pathway. As a result, this could create a loss of ventilation which may lead to inadequate oxygenation for patients, increasing the possibility of hypoxia.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new GE Healthcare, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 29, 2024 · Data as of Sep 19, 5:33 PM UTC