Reported Dec 12, 2012 · Terminated
NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number…
Philips And Neusoft Medical Systems Co., Ltd.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips And Neusoft Medical Systems Co., Ltd. · 16 Century Road, Neusoft Park, · Shenyang, Liaoning, N/A · N/A
Is this the product you have?
- NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.
- Lot codes and dates
- NeuViz 16 system Serial numbers
N16E090002N16E090003N16E090004N16E090007N16E090008N16E090010N16E100015N16E100017N16E100023N16E110016N16E110023N16E110026N16E110027N16E110040N16E110053N16E120016N16E120022N16E120020N16E120023N16E120024
- Neuviz Dual Serial numbers
400562400563400626NDH009EINDH011EINDH012EINDH013EINDH016EINDH017EINDH026EINDH027EINDH028EINDH030EINDH031EINDH034EINDHR080001NDHR080002NDHR080003NDHR080004NDHR080007NDHR080012NDHR080021NDHR090003NDHR090012and NDHR090013
- NeuViz 16 system Serial numbers
- Amount recalled
- 20 units of NeuViz 16 CT Scanner Systems and 25 units of NeuViz Dual series
Where it was sold
Why it was recalled
The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips And Neusoft Medical Systems Co., Ltd. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.