Reported Dec 5, 2018 · Terminated

Dimension Vista Lactate Dehydrogenase (LDI) Flex reagent cartridge, an in vitro diagnostic test for the quantitative…

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc. · 500 Gbc Dr Ms 514 · Newark, DE · 19702-2466

Is this the product you have?

  • Dimension Vista Lactate Dehydrogenase (LDI) Flex reagent cartridge, an in vitro diagnostic test for the quantitative measurement of lactate dehydrogenase in human serum and plasma on the Dimension Vista¿ System.Cat. No. K2054, Mat. No. 10464323
Lot codes and dates
    • 18240BA
    • 18199BA
    • 18162AC
    • 18155BB
    • 18106BC
    • 18073BE
    • 18036BC
    • 17342BA
Amount recalled
13435

Where it was sold

  • Nationwide

Why it was recalled

Siemens Healthcare Diagnostics has confirmed that Dimension Vista Lactate Dehydrogenase (LDI) may exhibit an erroneously elevated result when the preceding assay in the reaction cuvette is Lactic Acid (LA). The magnitude of the elevation may be variable. Both quality control and patient samples may be impacted. Incomplete removal of residual lactate dehydrogenase, a component of the LA reagent, has been identified as the cause of the potentially elevated result. A falsely elevated result for lactate dehydrogenase may lead to additional investigations to determine the etiology of tissue damage. Results would be correlated with clinical history and presentation in addition to other diagnostic laboratory testing.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 29, 2020 · Data as of Sep 19, 5:33 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.