Reported Nov 23, 2016 · Terminated
Advia Chemistry Calibrator For in vitro diagnostic use in the calibration of ADVIA¿ Chemistry systems for the following…
Siemens Healthcare Diagnostics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Healthcare Diagnostics, Inc. · 511 Benedict Ave · Tarrytown, NY · 10591-5005
Is this the product you have?
- Advia Chemistry Calibrator For in vitro diagnostic use in the calibration of ADVIA¿ Chemistry
- systems for the following methodsAlbumin (ALB, ALB_c), Direct Bilirubin (DBIL_2), Total Bilirubin (TBIL_2), Calcium (CA, CA_c, CA_2, CA_2c), Cholesterol (CHOL), Creatinine (CREA, CREA_2, CRE_2c, ECRE_2), Glucose Hexakinase (GLUH), Glucose Oxidase (GLUO), Iron (IRON_2), Lithium (LITH), Magnesium (MG, MG_c), Phosphorus - inorganic (IP, IP_c), Total Protein (TP, TP_c), Triglycerides (TRIG), Urea Nitrogen (UN, UN_c), Uric Acid (UA, UA_c)
- Lot codes and dates
- Device Listing # D033650 Model# SMN# 10312279
- Lot Numbers
680726680726A680726B680726C680726D
- Amount recalled
- 10,614 units
This recall covers 5 products
The agency filed one notice per product. You are reading one of them; the other 4 are below.
Where it was sold
- Nationwide
Why it was recalled
Siemens has confirmed that the DBIL_2 Assigned Calibrator Values (ACV) currently assigned to the affected lots may cause, on average, a positive proportional bias of approximately 41% on patient samples compared to the reference method for Direct Bilirubin measurement.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.