Reported Feb 16, 2022 · Ongoing
GE Healthcare MR superconducting magnets, a component of GE Healthcare Discovery MR750 3.0T System, nuclear magnetic…
GE Healthcare, LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: GE Healthcare, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615
Is this the product you have?
- GE Healthcare MR superconducting magnets, a component of GE Healthcare Discovery MR750 3.0T System, nuclear magnetic resonance imaging system.
- Lot codes and dates
- All systems.
- Amount recalled
- 17,228 devices total in this field correction
This recall covers 32 products
The agency filed one notice per product. You are reading one of them; the other 31 are below.
19 more are in the archive.
Where it was sold
Why it was recalled
The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
This product could cause temporary or reversible health problems — stop using it.