Reported Jan 8, 2014 · Terminated
Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, 311 Enterprise Dr Plainsboro,…
Integra LifeSciences Corp.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Integra LifeSciences Corp. · 311 Enterprise Dr · Plainsboro, NJ · 08536-3344
Is this the product you have?
- Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, 311 Enterprise Dr Plainsboro, NJ 08536 USA The Camino Advanced Monitor is a compact, portable device for use with Camino 110-4 series of Pressure / Temperature and Pressure Transducer-Tipped Catheters. Pressure and/or Temperature are measured at the Catheter tip, eliminating the need for external transducers, fluid, pressure tubing, and flush devices. The Camino displays intracranial pressure (ICP), intracranial temperature (ICT), and calculates cerebral perfusion pressure (CPP) (CPP = mean arterial pressure - mean intracranial pressure). The Camino provides a continuous display of the ICP waveform, as well as mean ICP, CPP, ICT, or systolic and diastolic waveform values. A continuous record of mean pressure and CPP values over the most recent 24-hour period is stored in memory, and can be displayed on command as a TREND either as the most recent 12 or 24 hour period. Outputs for ICP and ICT are available for use with patient bedside monitors. An isolated analogue output provides a continuous ICP waveform for hard copy documentation or data acquisition. An input to receive mean arterial pressure from a compatible patient bedside monitor is available for use when it is desired to have a CPP reading. The mean arterial pressure cannot be acquired from a non-invasive blood pressure cuff. Although the Camino is intended to be a stand-alone system, it also conveniently connects to hospital bedside monitoring systems.
- Lot codes and dates
275 of 275 codes
MEK1200104MEK1200204MEK1200304MEK1200404MEK1200504MEK1200604MEK1200704MEK1200804MEK1200904MEK1201004MEK1201104MEK1201204MEK1201304MEK1201404MEK1201504MEK1201604MEK1201704MEK1201804MEK1201904MEK1202004MEL1200104MEL1200204MEL1200304MEL1200404MEL1200504MEL1200604MEL1200704MEL1200804MEL1200904MEL1201004MEL1201104MEL1201204MEL1201304MEL1201404MEL1201504MEL1201604MEL1201704MEL1201804MEL1201904MEL1202004MEL1202104MEL1202204MEL1202304MEL1202404MEL1202504MEL1202604MEL1202704MEL1202804MEL1202904MEL1203004MEL1203104MEL1203204MEL1203304MEL1203404MEL1203504MEL1203604MEL1203704MEL1203804MEL1203904MEL1204004
- Amount recalled
- 275 CAM01 Monitors
Where it was sold
- Nationwide
Why it was recalled
Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is used on CAM01 monitor that contains Mod 4 Firmware, the displayed values for the systolic and diastolic intracranial pressure (ICP) are reversed. But, if the CAM01 is used with an auxiliary bedside monitor, the bedside monitor displays these values correctly.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Integra LifeSciences Corp. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.