Reported Jan 20, 2016 · Terminated

Perkin Elmer WIZARD2 10-detector, 1000 samples. Product Code: 2470-0200, 3470-0200. The Wizard gamma counter is…

Perkinelmer

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Perkinelmer · 940 Winter Street · Waltham, MA · 02451

Is this the product you have?

    • Perkin Elmer WIZARD2 10-detector
    • 1000 samples
  • Product Code2470-0200, 3470-0200. The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples
Lot codes and dates
  • Serial Numbes
    • DG08118243
    • DG12118732
    • DG09106871
    • DG03117554
    • DG05106373
    • DG02117481
    • DG11107153
    • DG07118106
    • DG03128981
    • DG04106316
    • DG03117574
    • DG04106296
    • DG04117674
    • DG05106414
    • DG02117455
    • DG07106607
    • DG11107048
    • DG03106163
    • DG07106604
    • DG05117805
    • DG08106759
    • DG03129008
Amount recalled
22 units

This recall covers 7 products

The agency filed one notice per product. You are reading one of them; the other 6 are below.

Where it was sold

  • Nationwide

Why it was recalled

The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new produce recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 4, 2017 · Data as of Sep 19, 6:02 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.