Reported Jan 20, 2016 · Terminated
Perkin Elmer WIZARD2 1-detector,3", 1000 samples. Product Code: 2480-0010 The Wizard gamma counter is intended to…
Perkinelmer
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Perkinelmer · 940 Winter Street · Waltham, MA · 02451
Is this the product you have?
Perkin Elmer WIZARD2 1-detector3"1000 samples
- Product Code2480-0010 The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples
- Lot codes and dates
- Serial NumbersDG04106256_REWORK
DG12107256DG04106256DG08118161DG03117633DG09118348DG06117938DG04106271DG03117625DG11118603DG03129003DG11107087DG09118269DG09118346DG09118323DG09118410DG10118462DG12118695DG03129030DG05117802DG04129089DG11107086DG03117631DG10118453DG01117421
163 of 163 codes
DG06106456DG03106146DG11118564DG03129000DG02106095DG08106711DG02128912DG10106940DG07118046DG07118051DG03117635DG09118273DG05117890DG12118657DG09118349DG03117557DG10118483DG10118478DG11107160DG04129156DG02106081DG09118167DG02128958DG03129032DG03117634DG11118575DG08118256DG05106334DG10106910DG10106950DG03117630DG05106409DG07106656DG12107176DG05106333DG05117810DG10106978DG03106122DG12107186DG08106662DG05117817DG05117874DG02128962DG11118610DG03129033DG03129034DG12118735DG11107053DG09118307DG07118070DG09118306DG08118170DG02128953DG02128936DG03129053DG08118162DG11118587DG03129031DG02128933DG03129002
- Serial Numbers
- Amount recalled
- 188 units
This recall covers 7 products
The agency filed one notice per product. You are reading one of them; the other 6 are below.
Where it was sold
- Nationwide
Why it was recalled
The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new produce recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.