Reported Mar 7, 2018 · Terminated

Footswitch accessory to the CUSA Clarity Ultrasonic Tissue Ablation System; Catalogue No: C7000 Product Usage: The…

Integra LifeSciences Corp.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Integra LifeSciences Corp. · 311 Enterprise Dr · Plainsboro, NJ · 08536-3344

Is this the product you have?

  • Footswitch accessory to the CUSA Clarity Ultrasonic Tissue Ablation System
  • Catalogue NoC7000
  • Product UsageThe footswitch is a component of the CUSA Clarity Ultrasonic Surgical Aspirator System, which is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue is desirable. The CUSA Clarity Ultrasonic Surgical
  • Aspirator is indicated for use inNeurosurgery, Plastic and Reconstructive surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific
  • uses are cleared for use in the US onlyGastrointestinal and Affiliated Organ Surgery including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy Urological surgery including removal of renal parenchyma during nephrectomy or partial nephrectomy General Surgery including removal of benign or malignant tumors or other unwanted soft tissue in open or minimally invasive general surgical procedures Laparoscopic Surgery including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy
Lot codes and dates
    • Lot #1885744
    • Lot #1923802
    • Lot #1985221
    • Lot #2057996
    • Lot #2109587
    • Lot #2118507
Amount recalled
155

Where it was sold

Why it was recalled

A faulty footswitch may result in unintended ultrasonic fragmentation activation. There are two scenarios when the failure mode may occur: 1) Ultrasonic fragmentation may be inadvertently activated at start-up or at the initial use once the footswitch is connected to the console and after the user primes the system, performs a handpiece test, and enters the Main Screen. 2) Ultrasonic fragmentation may remain activated during surgical use once the foot pedal is disengaged slowly.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Integra LifeSciences Corp. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 14, 2018 · Data as of Sep 19, 5:26 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.