Reported Jan 16, 2019 · Terminated
Ingenia Elition S, Ingenia Elition X
Philips Medical Systems Nederlands
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Medical Systems Nederlands · Veenpluis 4-6 · Best · N/A
Is this the product you have?
- Ingenia Elition S, Ingenia Elition X
- Lot codes and dates
- Software version R5.5.0
- Serial numbers
4502745011450104502445032450264505445035450514505245007450434504445031450394502546001450334505045034450494503745048460044504645008450584502845029450304504246005460114504045045
- Amount recalled
- 35
Where it was sold
Why it was recalled
During automatic tabletop movement, pressing and holding the Stop Table button may lead to an unexpected restart of the table movement.