Reported May 8, 2019 · Terminated

Ingenia Ambition S (781359), Ingenia Ambition X (781356) with Magnet Energization Device (MED), Philips MR systems are…

Philips Medical Systems Nederlands

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Philips Medical Systems Nederlands · Veenpluis 4-6 · Best, N/A · N/A

Is this the product you have?

  • Ingenia Ambition S (781359), Ingenia Ambition X (781356) with Magnet Energization Device (MED), Philips MR systems are indicated for use as a diagnostic device. The purpose of the MED is to supply power to charge the magnet. MED is abbreviation for Magnet Energize Device
Lot codes and dates
  • US Serial Number47004
  • OUS Serial Numbers
    • 47001
    • 47002
    • 48000
Amount recalled
4 units

Where it was sold

Why it was recalled

Magnet Energization Device (MED) can have a residual voltage on output terminal, may cause electrical shocks to personnel servicing the MED, MDD or connected cabling.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Philips Medical Systems Nederlands recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 4, 2020 · Data as of Sep 19, 4:52 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.