Reported Mar 6, 2013 · Terminated
GE Healthcare Definium 5000/5220493-2 X-Ray System
GE Healthcare, LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: GE Healthcare, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615
Is this the product you have?
- GE Healthcare Definium 5000/5220493-2 X-Ray System
- Lot codes and dates
- Definium Model 5000/5220493-2
- Amount recalled
- 312 installed in US
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
- Nationwide
Why it was recalled
When quickly switching the protocols from AEC to Fixed under the same view, if the user selection process time is less than 0.5 seconds, the User Interface (UI) may display different exposure techniques than selected, and may not display AEC setting. The exposure can still be made as commanded, however it may not be properly indicated.
This product could cause temporary or reversible health problems — stop using it.