Reported Jan 22, 2014 · Terminated

Draeger Infinity Acute Care System Monitoring Solution

Draeger Medical Systems, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Draeger Medical Systems, Inc. · 3135 Quarry Rd · Telford, PA · 18969-1042

Is this the product you have?

  • Draeger Infinity Acute Care System Monitoring Solution
Lot codes and dates
  • with software versions VG2.0.3 and higher.
Amount recalled
439

Where it was sold

Why it was recalled

After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.