Reported Feb 10, 2016 · Terminated

ADVIA Centaur PSA Assay; Catalog Number: 06574155/SMN 10310292 (100 test kit) and Catalog Number: 02676506/SMN 10310293…

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc · 333 Coney St · East Walpole, MA · 02032-1516

Is this the product you have?

  • ADVIA Centaur PSA Assay
  • Catalog Number06574155/SMN 10310292 (100
  • test kit) and Catalog Number02676506/SMN 10310293 (500
  • test kit) Product UsageThis in vitro diagnostic assay is intended to quantitatively measure prostate-specific Antigen (PSA) in human serum using the ADVIA Centaur/XP/XPT/CP Systems
Lot codes and dates
  • SMN 10310292-100 test kits Lot# UDI #
    • 54637263
    • 0063041420300354637263160223
    • 55304263
    • 0063041420300355304263160223
    • 56249264
    • 0063041420300356249264160327
    • 76727264
    • 0063041420300376727264160327
    • 00825265
    • 0063041420300300825265160424
    • 92416265
    • 0063041420300392416265160424
    • 99151265
    • 0063041420300399151265160424
    • 12970266
    • 0063041420300312970266160721
    • 14158266
    • 0063041420300314158266160721
    • 23016266
    • 0063041420300323016266160721
    • 28881266
    • 0063041420300328881266160721
    • 40776268
    • 0063041420300340776268160922
    SMN 10310293-500 test kits Lot# UDI #
    • 54448263
    • 0063041420299054448263160223
    • 55171263
    • 0063041420299055171263160223
    • 56171264
    • 0063041420299056171264160327
    • 77560264
    • 0063041420299077560264160327
    • 92415265
    • 0063041420299092415265160424
    • 96996265
    • 0063041420299096996265160424
    • 12971266
    • 0063041420299012971266160721
    • 13429266
    • 0063041420299013429266160721
    • 19651266
    • 0063041420299019651266160721
    • 27278266
    • 0063041420299027278266160721
    • 35089266
    • 0063041420299035089266160721
    • 39805268
    • 0063041420299039805268160922
    • 50851268
    • 0063041420299050851268160922
Amount recalled
55,754 kits

Where it was sold

Why it was recalled

PSA assay is not meeting the High Dose Hook Effect expectation in the Instructions for Use (IFU).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Nov 29, 2019 · Data as of Sep 19, 6:00 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.