Reported Jan 15, 2020 · Terminated

Cios Alpha, Model Number 10308191; mobile X-Ray system designed to provide X-ray imaging of the anatomical structures…

Siemens Medical Solutions USA, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418

Is this the product you have?

  • Cios Alpha, Model Number 10308191; mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical application
Lot codes and dates
  • Serial Numbers

    190 of 190 codes

    • 11294
    • 11202
    • 12009
    • 11141
    • 11163
    • 11032
    • 11311
    • 11260
    • 11258
    • 12530
    • 12334
    • 11376
    • 12426
    • 11336
    • 11132
    • 11295
    • 11044
    • 11005
    • 11333
    • 11386
    • 12597
    • 10076
    • 12677
    • 10115
    • 11291
    • 12523
    • 11056
    • 12319
    • 12623
    • 12512
    • 12574
    • 12647
    • 12014
    • 12452
    • 12306
    • 12317
    • 11119
    • 12576
    • 12678
    • 12513
    • 12335
    • 12490
    • 12657
    • 11200
    • 11281
    • 11283
    • 12425
    • 12023
    • 12396
    • 11285
    • 11205
    • 11103
    • 11105
    • 11257
    • 11074
    • 12379
    • 11013
    • 10021
    • 11330
    • 11066
  • Serial Numbers added as of 1/20/20
    • 10078
    • 10089
    • 11082
    • 11080
    • 11246
    • 11380
    • 11379
    • 12006
    • 12557
    • 12718
    • 12735
    • 12732
Amount recalled
190 (updated total of 202 as of 1/20/20)

Where it was sold

Why it was recalled

During a routine inspection of a system, a crack on the C-arm holder was discovered. In the event the C-arm holder has a crack, the stability of the C-arm may be compromised and could lead to deformations of the C-arm. In very unlikely cases, an extreme impact to the C-arm such as heavy collision with a wall or table, may lead to a detachment of the C-arm and could potentially result in injury to patients and staff.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Medical Solutions USA, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated May 6, 2020 · Data as of Sep 19, 5:26 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.