Reported Jan 29, 2014 · Terminated

ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems. Distributed by Philips Medical Systems, Cleveland,…

Philips Medical Systems (Cleveland) Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Philips Medical Systems (Cleveland) Inc · 595 Miner Rd · Cleveland, OH · 44143-2131

Is this the product you have?

  • ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems. Distributed by Philips Medical Systems, Cleveland, OH
Lot codes and dates
  • ADAC ARGUS

    161 of 161 codes

    • Model #882060
    • Serial #: 1023
    • 1178
    • 106785
    • 136560
    • 1100005
    • 9051005
    • 9304030
    • 9309011
    • 9501007
    • 9509005
    • 9510027
    • 9510028
    • 9603042
    • 9603043
    • 9604044
    • 9604047
    • 9605011
    • 9607060
    • 9609065
    • 9610072
    • 9704081
    • 9709090
    • 9710097
    • 9711103
    • 9712105
    • 9712106
    • 9801107
    • 9801109
    • 9802112
    • 9803001
    • 9803116
    • 9804126
    • 9804129
    • 9806136
    • 9807138
    • 9807422
    • 9810527
    • 9901156
    • 9901160
    • 9901162
    • 9908167
    • 9909172
    • 9910176
    • 00709B
    • 0101003A
    • 01310B
    • 01525B
    • 01920B
    • 01921B
    • 01922B
    • 03041B
    • 03094B
    • 03338B
    • 03368B
    • 2145-30074A
    • 7718B
    • A0103015
    • A0103213
    • A0105217
  • ADAC ARGUS
    • Model #88295
    • Serial #: 9211014
    • 9301005
    • 9304019
    • 9304024
    • 9304028
    • 9401028
    • 9401031
    • 9401037
    • 9505014
    • 9910033
    • 93-10678- 1292
    • 9910028P
    • M@10225AMT
    • M@3732B
    • M@3903B
    • M@3954B
    • M@3956B
    • M@3960B
    • M@4249B
    • M@4415B
    • M@4646B
    • M@4868B
    • M@5172B
    • M@5321B
    • M@7341B & M@AMT1154
  • ADAC CIRRUS
    • Model #88294
    • 34
    • 718558
    • 9206009
    • 9207003
    • 9207004
    • 9207005
    • 9207007
    • 9211006
    • 9300113
    • 9304013
    • 9309012
    • 9404022
    • 9407028
    • 9509042
    • 9604098
    • 9910030
    • 93-10735-0393
    • 9621RQ
    • M@10110AMT
    • M@10324AMT
    • M@10402AMT
    • M@3599G-08
    • M@3905Q
    • M@4029Q
    • M@4129Q
    • M@4142Q
    • M@4668Q
    • M@4743Q
    • M@5003Q
    • M@5668Q
    • M@5895Q
    • M@AMT126R
    • M@AMT1282 & M@MVS3227A
Amount recalled
255 units

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

The recalling firm determined that visual inspections of the fork joints behind the detector that are conducted during six month preventive maintenance (PM) may not detect small cracks or deficiencies in the fork weldment.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Philips Medical Systems (Cleveland) Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Nov 6, 2015 · Data as of Sep 19, 5:21 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.