Reported Feb 20, 2013 · Terminated
GE Healthcare, Dash 3000, Dash 4000 and Dash 5000 Physiological Patient Monitor and accessories.
GE Healthcare, LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: GE Healthcare, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615
Is this the product you have?
- GE Healthcare, Dash 3000, Dash 4000 and Dash 5000 Physiological Patient Monitor and accessories.
- Lot codes and dates
- FMI# &
- Descr36080 DASH 3/4/5k Custom Auto NBP System ID (As required)
2,169 of 2,169 codes
SD010210331GASD010262074GASD010262088GASD010272286GASD010272290GASD010272298GASD010272299GASD010272300GASD010272304GASD010272306GASD010272307GASD010272308GASD010272518GASD010272529GASD010272531GASD010272533GASD010272535GASD010272536GASD010272537GASD010272539GASD010272541GASD010272542GASD010272544GASD010272545GASD010272546GASD010272547GASD010292708GASD010292709GASD010292711GASD010292836GASD010292904GASD010292966GASD010292968GASD010292969GASD010292977GASD010292978GASD010292980GASD010292982GASD010292984GASD010292985GASD010292986GASD010292987GASD010292988GASD010292989GASD010292990GASD010292991GASD010292992GASD010292993GASD010292994GASD010292995GASD010292996GASD010292997GASD010292999GASD010293000GASD010293106GASD010293107GASD010293108GASD010293109GASD010293110GASD010293113GA
- Amount recalled
- 119,262
This recall covers 5 products
The agency filed one notice per product. You are reading one of them; the other 4 are below.
Where it was sold
Why it was recalled
GE has become aware of multiple issues affecting the Dash 3000/4000/5000 Patient Monitor. These issues may occur when the Ethernet controller in the XPC823 CPU on the Dash and DashPort 2 motherboard functions intermittently. GE Healthcare has identified that these issues occur most often at sites with large Unity Network installations. It should be noted that the Dash 3000/4000/5000 patient m
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new GE Healthcare, LLC recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.