Reported Mar 14, 2018 · Terminated
INFX-8000V Fluoroscopic X-Ray Systems; INFINIX CF-I/BP PLUS CARDIAC SYSTEM
Toshiba American Medical Systems Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Toshiba American Medical Systems Inc · PO Box 2068 · Tustin, CA · 92780-7047
Is this the product you have?
- INFX-8000V Fluoroscopic X-Ray Systems; INFINIX CF-I/BP PLUS CARDIAC SYSTEM
- Lot codes and dates
- Serial Numbers
UFD13X2015UFD13X2016UUC13Z2011UEC1422013UFD1442017UFD1452018AAA1642005UUC1452012UTC14Y2008ACA1472001ACA1492002AAA1492001AAA14X2002A8A1532001A9A1542001AAA1542003A9A1582002ABB1627001ACB1622003A9A1662003AAA1692006A9A16X2004A8B16Y2002A7B16Y2001ACB16Y2004AAA1632004AAA1712007W4B1622031W4B14Y2023
- Serial Numbers
- Amount recalled
- 29 units
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Why it was recalled
During an examination a device error occurred and the ceiling drive operation was disabled. It was found that the fastening bolts of the driving portion of the ceiling drive loosened cause the error.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Toshiba American Medical Systems Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.