Reported Mar 2, 2016 · Terminated
Philips IntelliVue Module Measurement X2 Model: M3002A
Philips Electronics North America Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Electronics North America Corporation · 3000 Minuteman Rd · Andover, MA · 01810-1032
Is this the product you have?
- Philips
- IntelliVue Module Measurement X2 ModelM3002A
- Lot codes and dates
- Software RevisionK.21.54 or L.00.96
- Amount recalled
- 13,300
This recall covers 5 products
The agency filed one notice per product. You are reading one of them; the other 4 are below.
Where it was sold
- Nationwide
Why it was recalled
The ST elevation alarm on the Patient Monitor or standalone X2 Measurement Module will not sound when indicated for all chest leads derived using Hexad 12-Lead ECG Monitoring in the Host Monitor
This product could cause temporary or reversible health problems — stop using it.