MiniMed 630G

Medtronic MiniMed

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Medtronic MiniMed · 18000 Devonshire St · Northridge, CA · 91325-1219

Is this the product you have?

MiniMed 630G

8 of 8 products

Products covered by this recall
#Product as published
1MMT-1715, MMT-1754, MMT-1755
2MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782
3MiniMed 770G: MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892
4MiniMed 620G (OUS Only): MMT-1750
5MiniMed 640G (OUS Only): MMT-1711, MMT-1712, MMT-1751, MMT-1752
6MiniMed 720G (OUS Only): MMT-1809, MMT-1810, MMT-1859, MMT-1860
7MiniMed 740G (OUS Only): MMT-1811, MMT-1812, MMT-1861, MMT-1862
8MiniMed 780G (OUS Only): MMT-1884, MMT-1885, MMT-1886, MMT-1895, MMT-1896 The MiniMed" 630G System with SmartGuard is intended for continuous delivery of basal insulin (at user selected rates) and administration of insulin boluses (in user selectable amounts) for the management of Type 1 diabetes mellitus in persons sixteen years of age and older.
Lot codes and dates
  • Insulin pumps distributed in the last 6 months. All devices (US) serial numbers ranging from NG2056884H TO NG2891033H and (OUS) serial numbers ranging from NG1921788H TO NG2883091
Amount recalled
175,801

Where it was sold

Why it was recalled

New or replacement insulin pumps are not pre-programmed with basal rates or other verified settings (bolus wizard settings, sensor settings, etc.), which must be set up and saved on pumps prior to use. Patients who do not realize this will receive no basal insulin, which can lead to insulin under-delivery. Insulin under-delivery can cause severe hyperglycemia, which may lead to diabetic ketoacidosis. Current device labeling does not clearly identify that these insulin pumps are capable of functioning with the default basal setting of 0.0 units/hr for all 24 hours and that no error message will appear if the default basal setting remains active. Furthermore, the labeling for the affected devices does not highlight that saving the basal settings requires multiple steps and failure to complete all steps will not save the new basal settings, nor that the steps required to set and save a basal profile and other parameters which are used as part of the bolus calculator function are different.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Medtronic MiniMed recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:37 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.