MiniMed 780G Insulin Pump: Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL;¿ PUMP MMT-1886 MM780G 6.5W BLE MG; KIT…
Medtronic Minimed
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Medtronic Minimed · 18000 Devonshire St · Northridge, CA · 91325-1219
Is this the product you have?
- MiniMed 780G Insulin Pump
Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL¿ PUMP MMT-1886 MM780G 6.5W BLE MGKIT MMT-1895WW MM780G 6.5V BLE SF MMOL¿ KIT MMT-1896WW MM780G 6.5W BLE SF MG¿ PUMP MMT-1884XC 780G V6.5 CLIN US MG¿ PUMP MMT-1885XC 780G V6.5 CLIN MM¿ PUMP MMT-1886XC 780G V6.5 CLIN MG¿ PUMP MMT-1885L MM780G 6.5V BLE MMOL¿ PUMP MMT-1886L MM780G 6.5W BLE MG¿ KIT MMT-1896ES MM780G V6.5W MG¿ KIT MMT-1896WWA MM780G BLE MG ES¿ KIT MMT-1895WWA MM780G BLE MMOL AR¿ KIT MMT-1895WWA MM780G BLE MMOL CS¿ KIT MMT-1895WWA MM780G BLE MMOL DA¿ KIT MMT-1895WWA MM780G BLE MMOL DE¿ KIT MMT-1895WWA MM780G BLE MMOL EN¿ KIT MMT-1895WWA MM780G BLE MMOL FI¿ KIT MMT-1895WWA MM780G BLE MMOL FR¿ KIT MMT-1895WWA MM780G BLE MMOL IT¿ KIT MMT-1895WWA MM780G BLE MMOL NL¿ KIT MMT-1895WWA MM780G BLE MMOL NO¿ KIT MMT-1895WWA MM780G BLE MMOL SLKIT MMT-1895WWA MM780G BLE MMOL SV¿ KIT MMT-1895WWA MM780G BLE MMOL HU¿ KIT MMT-1896WWA MM780G BLE MG AR¿ KIT MMT-1896WWA MM780G BLE MG DE¿ KIT MMT-1896WWA MM780G BLE MG EL¿ KIT MMT-1896WWA MM780G BLE MG HE¿ KIT MMT-1896WWA MM780G BLE MG IT¿ KIT MMT-1896WWA MM780G BLE MG PL¿ KIT MMT-1896WWA MM780G BLE MG RO¿ KIT MMT-1896WWA MM780G BLE MG TR¿ KIT MMT-1896WWA MM780G BLE MG NL¿ KIT MMT-1896WWA MM780G BLE MG FR¿ KIT MMT-1896WWA MM780G BLE MG ES
- Lot codes and dates
2,453 of 2,453 products
Products covered by this recall # Barcodes Product as published 1 00763000384258MMT-1885/00763000384258 2 00763000384265¿ MMT-1886/00763000384265 3 00763000384272¿ MMT-1895WW/00763000384272 4 00763000384289¿ MMT-1896WW/00763000384289 5 00763000455613¿ MMT-1884XC/ 00763000455613 6 00763000421939¿ MMT-1885XC/00763000421939 7 00763000421946¿ MMT-1886XC/00763000421946 8 00763000441432¿ MMT-1885L/00763000441432 9 00763000441449¿ MMT-1886L/00763000441449 10 00763000454296¿ MMT-1896ES/00763000454296 11 00763000411541¿ MMT-1896WWA/00763000411541 12 00763000411374¿ KIT MMT-1895WWA/00763000411374 13 00763000411381¿ KIT MMT-1895WWA/00763000411381 14 00763000411398¿ KIT MMT-1895WWA/00763000411398 15 00763000411404¿ KIT MMT-1895WWA/00763000411404 16 00763000411411¿ KIT MMT-1895WWA/00763000411411 17 00763000411428¿ KIT MMT-1895WWA/00763000411428 18 00763000411435¿ KIT MMT-1895WWA/00763000411435 19 00763000411459¿ KIT MMT-1895WWA/00763000411459 20 00763000411466¿ KIT MMT-1895WWA/00763000411466 - Amount recalled
- O.U.S. Distibution only = 15,682 devices
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
Due to a software design issue, under certain conditions, a software fault is detected when a large bolus delivery at quick bolus speed completes. if the user is not aware of the amount of active insulin and delivers an additional bolus, there is a risk of insulin over delivery.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Medtronic Minimed recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.