Reported Jan 20, 2021 · Ongoing

Discovery NM 630 (Model Number H3101RH), NM 830 (Model Number H3910AC), 830 NM Cyber Defense (Model Number H3910AG) -…

GE Healthcare, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE Healthcare, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615

Is this the product you have?

    • Discovery NM 630 (Model Number H3101RH)
    • NM 830 (Model Number H3910AC)
    • 830 NM Cyber Defense (Model
  • Number H3910AG) - Product Usageintended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body
Lot codes and dates
  • Discovery NM 630 (Model Number H3101RH)
  • Serial numbers
    • 63GY50110
    • 63GY50119
    • 63GY50116
    • 63GZ50141
    • 63GY50120
    • 63GY50108
    • 63GY50121
    • 63GZ50142
    • 63GY50112
    • 63GZ50139
    • 63GY50107
    • 63GZ50138 NM 830 (Model Number H3910AC)
  • Serial Numbers830Z60199, 830Z60182, 830Z60183, 830Z60192, 830Y60089, 830Y60105, 830Y60079, 830Z60207, 830Y60092, 830Y60102, 830Y60081, 830Z60200, 830Y60106, 830Z60201, 83EZ88001, 830Y60111, 830Y60094, 830Z60211, 830Y60101, 830Z60210, 830Y60118, 830Z60195, 830Z60194, 830Z60198, 830Z60181, 830Z60202, 830Y60097, 830Z60196, 830Z60180, 830Y60078, 830Y60115, 830Y60114, 830Y60108, 830Y60100, 830Y60087, 830Y60084, 830Z60188, 830Z60206, 830Y60112, 830Y60082, 830Y60093, 830Z60191, 830Y60107, 830Y60080, 830Z60190, 830Y60103, 830Z60193, 830Y60095, 830Y60119, 830Z60209, 830Z60205, 830Y60073, 830Y60120, 830Y60083, 830Y60116, 830Z60203, 830Y60085, 830Y60077, 830Z60184, 830Y60086, 830Z60187, 830Y60113, 830Z60208, 830Y60110, 830Z60204, 830Y60072, 830Z60179, 830Z60186 830 NM Cyber Defense (Model Number H3910AG)
  • Serial Numbers830Z60185
Amount recalled
81 units

This recall covers 6 products

The agency filed one notice per product. You are reading one of them; the other 5 are below.

Why it was recalled

GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new GE Healthcare, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:57 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.