Reported Feb 7, 2024 · Ongoing
smiths medical Medfusion Model 3500 Syringe pump
Smiths Medical ASD Inc.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Smiths Medical ASD Inc. · 6000 Nathan Ln N · Minneapolis, MN · 55442-1690
Is this the product you have?
- smiths medical Medfusion Model 3500 Syringe pump
- Lot codes and dates
- All Medfusion Model 3500 pumps with software v3.X, v4.X, and v5.0.0 and was addressed in software v6.0.0 (2012) and all subsequent versions.
- Amount recalled
- 93177 pumps
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Why it was recalled
Stop using this product now
Medfusion syringe pumps, with software versions before v6.0.0, may have the following issues:1. Delivery During Motor Not Running High Priority Alarm, 2. Infusion Restarted with Incorrect Parameters, 3. Screen Lock, 4. Interruption of Bolus or Loading Dose Delivery, 5. Pump Displays Incorrect Bolus/Loading Dose, 6. Loading/Bolus Dose Below the Minimum Recommended Rate, 7. Motor Rate Error, 8. Incorrect Recall Last Settings, 9. Corrupt Configuration, 10. Auto Lock, 11. Toolbox Configuration Loading Dose Time Values. Smiths Medical corrected all issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent Medfusion software (v6.0.0) installed on your pumps.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Smiths Medical ASD Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.