Reported Dec 17, 2025 · Ongoing
Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,
Bard Peripheral Vascular Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Bard Peripheral Vascular Inc · 1625 W 3rd St Bldg 1 · Tempe, AZ · 85281-2438
Is this the product you have?
- Safe-T-Centesis 6 Fr Catheter Drainage Tray
- REFPIG1260T
- Lot codes and dates
- UDI/Lot/Expiration(01)10885403108365(17)271130(10)0001623320/0001623320/20271130; (01)10885403108365(17)271130 (10)0001623069/0001623069/20271130
- Amount recalled
- 1240
Where it was sold
Why it was recalled
Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Bard Peripheral Vascular Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.