Reported Oct 8, 2025 · Ongoing

Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the…

Bard Peripheral Vascular Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Bard Peripheral Vascular Inc · 1625 W 3rd St Bldg 1 · Tempe, AZ · 85281-2438

Is this the product you have?

  • Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically configures non-adjustable treatment parameters (including time and temperature) for compatible catheters connected via a triaxial connector port. Designed for use with Venclose RF Catheters (either the EVSRF or Maven variants) as an integrated system, the device employs resistive radiofrequency ablation to deliver temperature-controlled RF energy that heats incompetent vein walls, inducing irreversible luminal occlusion followed by progressive fibrosis and eventual vein resorption. The 510(k) submission K250068 encompasses solely the Venclose digiRF Generator due to software modifications with no alterations to the cleared Venclose Catheters (EVSRF and Maven variants, cleared under K160754 and K211806, respectively).
Lot codes and dates
  • UDI-DI00858254006022. Generator Software Version 3.35
Amount recalled
1,725

Where it was sold

Why it was recalled

Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification process, the generator displays a Red X indicator on the screen without accompanying error codes, effectively disabling catheter functionality. However, BD has determined that software version 3.35 generates false positive failures due to temperature-dependent verification parameters, incorrectly flagging properly functioning Venclose EVSRF Ablation Catheters as defective when the catheter temperature is not at steady state during the check initiation.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Bard Peripheral Vascular Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:58 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.