Reported Mar 2, 2016 · Terminated

Mammomat Inspiration with Tomosynthesis functionality

Siemens Medical Solutions USA, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418

Is this the product you have?

  • Mammomat Inspiration with Tomosynthesis functionality
Lot codes and dates
  • Model number10140000
Amount recalled
76 units

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

Why it was recalled

A potential system fatal error may occur during tomosynthesis reconstruction with large breasts that cover nearly the entire detector surface and with a breast thickness larger than 90 mm. The data to be processed for the tomosynthesis application can exceed the capacity of the graphical processing unit (GPU). If the fatal error message occurs, reconstruction of the data will be aborted. The syste

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Medical Solutions USA, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Nov 8, 2017 · Data as of Sep 19, 6:01 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.