Cardiohelp-i Transport Guard with Transport Guard (701072525): is a blood oxygenation and carbon dioxide removal system…

Maquet Medical Systems USA

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Maquet Medical Systems USA · 45 Barbour Pond Dr · Wayne, NJ · 07470-2094

Is this the product you have?

  • Cardiohelp-i Transport
  • Guard with Transport Guard (701072525)is a blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support for periods appropriate to cardiopulmonary bypass (up to six hours)
  • Product Code/REF Number701072780 (Cardiohelp-i)
Lot codes and dates
  • Serial Numbers for the Cardiohelp-i (Part Number 701072780)as the Transport Guard (701072525) is packaged with the Cardiohelp-i and is not
  • labelled individually Serial Numbers
    • 90413945
    • 90414225
    • 90414226
    • 90414227
    • 90414228
    • 90414229
    • 90414232
    • 90414233
    • 90414234
    • 90414235
    • 90414236
    • 90414237
    • 90414283
    • 90414284
    • 90414285
    • 90414286
    • 90414287
    • 90414288
    • 90414289
    • 90414292
    • 90414293
    • 90414294
    • 90414295
    • 90414296
    • 90414297
    • 90414309
    • 90414310
    • 90414311
    • 90414312
    • 90414317
    • 90414318
    • 90414319
  • UDI04058863074863 (Cardiohelp-i)
Amount recalled
US: 32 units; OUS 30 units

Where it was sold

Why it was recalled

Potential for a loose cylindrical pin on the Transport Guard that may result in the jamming of the locking mechanism causing the Transport Guard to not open or, if the locking pin is fully disengaged from the assembly, a full disassembly of the Transport Guard may occur, resulting in a delay of support if a product exchange is necessary.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Maquet Medical Systems USA recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:33 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.