Reported Mar 21, 2018 · Terminated
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6) (GTIN 00085412091570); Model Numbers: 35700BAX…
Baxter Healthcare Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Baxter Healthcare Corporation · 711 Park Ave · Medina, NY · 14103-1036
Is this the product you have?
- Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6) (GTIN 00085412091570)
- Model Numbers35700BAX and 35700ABB. Intended to be used for the controlled administration of fluids
- Lot codes and dates
- 42
4,613 of 4,613 codes
700000700001700008400012700014700015700038400040400044400057700060400068700070400073400076700078700079400085700088700099700105400106400114700117700131700137400141700148700180700194700197400226700227700228700229700232700236700239400245700252700254700260700262700274400282700291700300700305700306400311700313700336700347700351700352400355700356700358400368700374
- Amount recalled
- 173,375
Where it was sold
Why it was recalled
Certain infusion pumps may exhibit excessive wear of the pumping mechanism caused by a lack of lubrication, this could affect the occlusion of the tubing and result in over- or under-infusion.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Baxter Healthcare Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.