Reported Mar 13, 2019 · Terminated
ARCHITECT MAGNESIUM, (a) REF 7D70-21, (b) 7D70-31 The Magnesium assay is used for the quantitation of magnesium in…
Abbott Laboratories, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Abbott Laboratories, Inc · 1921 Hurd Dr · Irving, TX · 75038-4313
Is this the product you have?
- ARCHITECT MAGNESIUM, (a) REF 7D70-21, (b) 7D70-31 The Magnesium assay is used for the quantitation of magnesium in human serum, plasma, or urine.
- Lot codes and dates
- (a) 7D70 21 Lot Number UDI Number 36338UN17 (01)00380740012922(17)190106(10)36338UN17 79865UN17 (01)00380740012922(17)190309(10)79865UN17 12921UN17 (01)00380740012922(17)190503(10)12921UN17 43618UN17 (01)00380740012922(17)190609(10)43618UN17 45337UN17 (01)00380740012922(17)190714(10)45337UN17 84994UN17 (01)00380740012922(17)191012(10)84994UN17 24856UN17 (01)00380740012922(17)191113(10)24856UN17 41652UN18 (01)00380740012922(17)200120(10)41652UN18 99639UN18 (01)00380740012922(17)200226(10)99639UN18 99661UN18 (01)00380740012922(17)200413(10)99661UN18 78493UN18 (01)00380740012922(17)200725(10)78493UN18 95220UN18 (01)00380740012922(17)200907(10)95220UN18 31428UN18 (01)00380740012922(17)201031(10)31428UN18 33784UN18 (01)00380740012922(17)200611(10)33784UN18 (b) 7D70-31 36339UN17 (01)00380740012939(17)190106(10)36339UN17 97389UN17 (01)00380740012939(17)190309(10)97389UN17 12922UN17 (01)00380740012939(17)190503(10)12922UN17 43616UN17 (01)00380740012939(17)190609(10)43616UN17 45338UN17 (01)00380740012939(17)190714(10)45338UN17 84993UN17 (01)00380740012939(17)190901(10)84993UN17 08788UN17 (01)00380740012939(17)191012(10)08788UN17 24857UN17 (01)00380740012939(17)191113(10)24857UN17 41651UN18 (01)00380740012939(17)200129(10)41651UN18 99640UN18 (01)00380740012939(17)200226(10)99640UN18 99662UN18 (01)00380740012939(17)200413(10)99662UN18 33783UN18 (01)00380740012939(17)200509(10)33783UN18 33785UN18 (01)00380740012939(17)200611(10)33785UN18 78859UN18 (01)00380740012939(17)200725(10)78859UN18 95341UN18 (01)00380740012939(17)200907(10)95341UN18 31429UN18 (01)00380740012939(17)201031(10)31429UN18
- Amount recalled
- 24,995 units
Where it was sold
- Nationwide
Why it was recalled
Abbott internal testing has identified that the Magnesium urine application demonstrates depressed urine result recovery. All samples >0.5mEq/L fail the linearity acceptance criteria for the bias specification and demonstrate depressed recovery of up to 37%. The specific cause of the negative bias/depressed results is currently under investigation.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Abbott Laboratories, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.