Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation…
Philips North America
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips North America · 222 Jacobs St · Cambridge, MA · 02141-2296
Is this the product you have?
- Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
- Lot codes and dates
- UDI-DI 00884838088115 Ingenia Serial number / Accessory Serial Number 45294 100029 45294 100030
- Amount recalled
- 2
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
Why it was recalled
Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.