Reported May 4, 2022 · Ongoing

IMMULITE 2000 Systems HCG [200 Tests, material number 10381206, catalog number L2KCG] and [600 Tests, material number…

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc. · 511 Benedict Ave · Tarrytown, NY · 10591-5005

Is this the product you have?

    • IMMULITE 2000 Systems HCG [200 Tests
    • material number 10381206
    • catalog number L2KCG] and [600 Tests
    • material number 10381194
    • catalog number L2KCG6]
Lot codes and dates
  • 8
  • production lots
    • 463
    • 464
    • 466
    • 467
    • 469
    • 470
    • 471
    • 472
    • [UDI
  • for 200 tests
    • (01)00630414961132(10)463(17)20220228
    • (01)00630414961132(10)464(17)20220331
    • (01)00630414961132(10)466(17)20220331
    • (01)00630414961132(10)467(17)20220531
    • (01)00630414961132(10)469(17)20220731
    • (01)00630414961132(10)470(17)20220930
    • (01)00630414961132(10)471(17)20221130
    • (01)00630414961132(10)472(17)20230228] [UDI
  • for 600 tests
    • (01)00630414961149(10)463(17)20220228
    • 01)00630414961149(10)464(17)20220331
    • (01)00630414961149(10)466(17)20220331
    • (01)00630414961149(10)467(17)20220531
    • (01)00630414961149(10)469(17)20220731
    • (01)00630414961149(10)470(17)20220930
    • (01)00630414961149(10)471(17)20221130
    • (01)00630414961149(10)472(17)20230228]
Amount recalled
10,755 boxes (533 US, 10,222 OUS)

Where it was sold

Why it was recalled

Potential for falsely elevated hCG results due to sample carryover in serum and urine patient samples, as well as quality control samples and adjustors.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:26 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.