Reported May 11, 2022 · Ongoing
ADVIA Centaur Enhanced Estradiol (eE2) 100T - In vitro diagnostic use in the quantitative determination of estradiol…
Siemens Healthcare Diagnostics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Healthcare Diagnostics, Inc. · 333 Coney St · East Walpole, MA · 02032-1516
Is this the product you have?
- ADVIA Centaur Enhanced Estradiol (eE2) 100T - In vitro diagnostic use in the quantitative
- determination of estradiol SMN10490889 All lots
- Lot codes and dates
- All lot numbers UDI(01)00630414573342(10)08930076(17)20220115 (01)00630414573342(10)10982085(17)20221108 (01)00630414573342(10)19929072(17)20210819 (01)00630414573342(10)20298079(17)20220304 (01)00630414573342(10)48171081(17)20220304 (01)00630414573342(10)53554081(17)20220521 (01)00630414573342(10)70551083(17)20220805 (01)00630414573342(10)76055076(17)20220115
- Amount recalled
- 2,315 units
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
- Nationwide
- GA
Why it was recalled
Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.