Reported Mar 21, 2018 · Terminated
CR POR FEM HATCP SIZE E/L SIZE F/L Knee Prosthesis
Zimmer Biomet, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Zimmer Biomet, Inc. · 1800 W Center St · Warsaw, IN · 46580-2304
Is this the product you have?
- CR POR FEM HATCP SIZE E/L SIZE F/L Knee Prosthesis
- Lot codes and dates
- Item Number65-5972-015-01
- Lot Number62086845
- Item Number65-5972-016-01
- Lot Numbers11007043 62105229
- Amount recalled
- 13,227 in total
This recall covers 59 products
The agency filed one notice per product. You are reading one of them; the other 58 are below.
46 more are in the archive.
Where it was sold
- Nationwide
Why it was recalled
The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.