Reported Mar 27, 2019 · Terminated
PET Plates Coated Film-MONOMEDBV2S2- Radiographic Film REF: XJUB8, XJM73, XJUDD, XJM85, EKLYZ. Product Usage: Due to an…
Agfa N.V.
Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Agfa N.V. · Septestraat 27 · Mortsel, N/A · N/A
Is this the product you have?
- PET Plates Coated Film-MONOMEDBV2S2-
- Radiographic Film REF
XJUB8XJM73XJUDDXJM85EKLYZ
- Product UsageDue to an inhomogeneous coating solution, pinholes can become visible in even dark images. Visibility of the pinholes varies as material from this coating run shows pinholes in different frequencies and intensities
- Lot codes and dates
REF CODE/ BATCH XJUB8 36930048XJM73 36930048XJUDD 36930047XJUDD 36940001XJM85 36940001XJUDD 36940002EKLYZ 36940003
- Amount recalled
- 2029 boxes
Where it was sold
- Nationwide
- OH
Why it was recalled
Due to an inhomogeneous coating solution, pinholes can become visible in even dark images. Visibility of the pinholes varies as material from this coating run shows pinholes in different frequencies and intensities
This product is unlikely to cause harm, but it breaks a safety or labelling rule.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Agfa N.V. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.