Reported Mar 25, 2020 · Terminated

AGFA Digital Radiography X- Ray system DR 800 with TomoSynthesis- a Tomographic X-ray system - Product Usage:…

Agfa N.V.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Agfa N.V. · Septestraat 27 · Mortsel · N/A

Is this the product you have?

  • AGFA Digital Radiography X- Ray system DR 800 with TomoSynthesis- a Tomographic
  • X-ray system - Product UsageTomoSynthesis is used to synthesize tomographic slices from a single tomographic sweep. The DR 800 is not intended for mammography applications
Lot codes and dates
  • Serial Numbers
    • 00949
    • 1161
Amount recalled
2 units

Where it was sold

Why it was recalled

Tomosynthesis image acquisition sequence did not stop automatically after expected number of exposures.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Agfa N.V. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 30, 2023 · Data as of Sep 19, 6:13 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.