Reported Apr 17, 2013 · Terminated

Giraffe OmniBed¿ and Giraffe Incubator p/n 6600-0723-700 (rev A and B) labeled in part: ***GE Healthcare 8880 Gorman…

GE Healthcare, LLC

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE Healthcare, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615

Is this the product you have?

  • Giraffe OmniBed¿ and Giraffe Incubator p/n 6600-0723-700 (rev
  • A and B) labeled in part***GE Healthcare 8880 Gorman Road Laurel, MD 20723 U.S.A. www.gehealthcare.com***. GE Healthcare Giraffe Omnibed is a combination of an infant incubator and an infant warmer
Lot codes and dates
  • OmniBed Serial Numbers

    464 of 464 codes

    • HDGR52546
    • HDGR52736
    • HDGR52802
    • HDGR52861
    • HDGR52910
    • HDGR52958
    • HDGR53007
    • HDGR53058
    • HDGR53110
    • HDGR53185
    • HDGR52547
    • HDGR52737
    • HDGR52803
    • HDGR52862
    • HDGR52911
    • HDGR52959
    • HDGR53008
    • HDGR53059
    • HDGR53112
    • HDGR53186
    • HDGR52558
    • HDGR52739
    • HDGR52804
    • HDGR52863
    • HDGR52912
    • HDGR52960
    • HDGR53009
    • HDGR53060
    • HDGR53113
    • HDGR53187
    • HDGR52609
    • HDGR52740
    • HDGR52805
    • HDGR52864
    • HDGR52913
    • HDGR52961
    • HDGR53010
    • HDGR53061
    • HDGR53114
    • HDGR53188
    • HDGR52610
    • HDGR52741
    • HDGR52806
    • HDGR52865
    • HDGR52914
    • HDGR52962
    • HDGR53011
    • HDGR53062
    • HDGR53115
    • HDGR53189
    • HDGR52613
    • HDGR52742
    • HDGR52813
    • HDGR52866
    • HDGR52915
    • HDGR52963
    • HDGR53012
    • HDGR53063
    • HDGR53116
    • HDGR53190
  • Serial Numbers

    215 of 215 codes

    • HDHR52549
    • HDHR52708
    • HDHR52774
    • HDHR52881
    • HDHR53029
    • HDHR52551
    • HDHR52709
    • HDHR52830
    • HDHR52882
    • HDHR53030
    • HDHR52552
    • HDHR52710
    • HDHR52831
    • HDHR52883
    • HDHR53031
    • HDHR52553
    • HDHR52711
    • HDHR52832
    • HDHR52884
    • HDHR53032
    • HDHR52554
    • HDHR52720
    • HDHR52833
    • HDHR52885
    • HDHR53039
    • HDHR52555
    • HDHR52721
    • HDHR52834
    • HDHR52894
    • HDHR53040
    • HDHR52556
    • HDHR52722
    • HDHR52835
    • HDHR52900
    • HDHR53041
    • HDHR52557
    • HDHR52723
    • HDHR52836
    • HDHR52901
    • HDHR53042
    • HDHR52559
    • HDHR52724
    • HDHR52837
    • HDHR52902
    • HDHR53043
    • HDHR52561
    • HDHR52725
    • HDHR52838
    • HDHR52903
    • HDHR53044
    • HDHR52562
    • HDHR52726
    • HDHR52839
    • HDHR52904
    • HDHR53045
    • HDHR52563
    • HDHR52727
    • HDHR52840
    • HDHR52905
    • HDHR53046
Amount recalled
679 units

This recall covers 5 products

The agency filed one notice per product. You are reading one of them; the other 4 are below.

Where it was sold

Why it was recalled

Stop using this product now

Neonatal beds and incubator display boards may randomly overwrite stored information and radiant heater and oxygen set points upon power up or recovery from a power failure.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new GE Healthcare, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jul 10, 2013 · Data as of Sep 19, 6:54 PM UTC