Reported Feb 1, 2017 · Terminated

Brilliance iCT

Philips Medical Systems (Cleveland) Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Philips Medical Systems (Cleveland) Inc · 595 Miner Rd · Cleveland, OH · 44143-2131

Is this the product you have?

  • Brilliance iCT
Lot codes and dates
  • Rotor-4535-673-94382 SN 19, installed on Brilliance ICT System S/N 200203
Amount recalled
1-unit

Why it was recalled

During a retrospective review of production documentation it was observed that the M6 Rotor Counterweight Bolt may have been improperly torqued.

This product could cause temporary or reversible health problems — stop using it.