Reported May 25, 2022 · Ongoing
MicroScan WalkAway-40 plus Reconditioned Instrument REF B1018-340R
Beckman Coulter Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Beckman Coulter Inc. · 250 S Kraemer Blvd · Brea, CA · 92821-6232
Is this the product you have?
- MicroScan WalkAway-40 plus Reconditioned Instrument REF B1018-340R
- Lot codes and dates
- Catalog NumberB1018-340R
- UDI-DI Code15099590684112
- Serial Numbers34020035 34020224 34030054
- Amount recalled
- 3 systems
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
Why it was recalled
WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
This product could cause temporary or reversible health problems — stop using it.