Reported Feb 8, 2017 · Terminated
IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LKUP1, SMN 10380956 (OUS Kit Catalog) Product…
Siemens Healthcare Diagnostics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Healthcare Diagnostics, Inc. · 511 Benedict Ave · Tarrytown, NY · 10591-5005
Is this the product you have?
IMMULITE/IMMULITE 1000 SystemsThird Generation PSACatalog Number/REF LKUP1SMN 10380956 (OUS
- Kit Catalog) Product UsageFor in vitro diagnostic use with the IMMULITE¿ and IMMULITE 1000 Analyzers for the quantitative measurement of prostate-specific antigen (PSA) in human serum, as an aid in the detection of prostate cancer when used in conjunction with digital rectal examination (DRE) in men aged 50 years or older. This assay is further indicated as an adjunctive test to aid in the management of prostate cancer patients
- Lot codes and dates
- Lot Numbers1) 0336 - Expiration date 2017-01-31 2) 0337 - Expiration date 2017-02-28 3) 0338 - Expiration date 2017-05-31
- Amount recalled
- 1379 kits (190 kits in U.S.A and 1189 Kits internationally) in total
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
Why it was recalled
Negative bias with IMMULITE¿ /IMMULITE¿ 1000 Third Generation PSA (prostate-specific antigen) kit. The investigation by Siemens has confirmed that patient samples exhibit an average negative bias of approximately -22% with a range of -16% to -27% for samples recovering from 0.08 to 16.2 ng/mL.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.