Reported May 8, 2013 · Terminated
GE, VCT for Discovery VCT, Discovery RX VCT, and Discovery PET/CT 690, Lightspeed VCT, Discovery CT750HD. Intended to…
GE Healthcare, LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: GE Healthcare, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615
Is this the product you have?
- GE, VCT for Discovery VCT, Discovery RX VCT, and Discovery PET/CT 690, Lightspeed VCT, Discovery CT750HD. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission.
- Lot codes and dates
- Serial Number
166 of 166 codes
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
- Serial Number
- Amount recalled
- 166
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
Why it was recalled
Fastener bolts on the heat exchanger were not properly torqued and part of the heat exchanger can detach from the gantry and hit the inside of the covers and other components of the gantry possibly resulting in the oil conduit breaking, spilling oil within the gantry and outside the system. This issue affects GE Healthcare VCT for Discovery VCT, Discovery RX VCT, and Discovery PET/CT 690, Lightspe
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new GE Healthcare, LLC recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.