Reported Mar 12, 2025 · Ongoing
Paradigm insulin pump, REF: MMT-754
Medtronic MiniMed, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Medtronic MiniMed, Inc. · 18000 Devonshire St · Northridge, CA · 91325-1219
Is this the product you have?
- Paradigm insulin pump
- REFMMT-754
- Lot codes and dates
- All Lot Numbers; All UDI numbers
- Amount recalled
- 41,404 units
This recall covers 12 products
The agency filed one notice per product. You are reading one of them; the other 11 are below.
Where it was sold
Why it was recalled
All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Medtronic MiniMed, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.